How the sheet works out pass or fail
For each test point the sheet subtracts the nominal value from your reading. That difference is the error. It then compares the size of the error with the tolerance at that point:
- Percent of span: tolerance × (range maximum − range minimum) ÷ 100. The limit is the same at every point.
- Percent of reading: tolerance × nominal value ÷ 100. The limit grows with the reading.
- Absolute: a fixed amount in the unit, such as ±0.2 °C.
- Percent of reading plus floor: the percent-of-reading limit plus a fixed amount, as many multimeter specifications are written.
A point passes when the error is no larger than the limit. A 0 to 100 psi gauge with a tolerance of ±0.5 % of span has a limit of ±0.5 psi, so a reading of 50.9 psi at 50.0 psi fails and a reading of 50.1 psi passes.
Recording as-found and as-left readings
- Take the as-found readings before you touch any adjustment. They are the evidence of how the instrument performed in service.
- If every point is within tolerance and you make no adjustment, the as-left readings are the same. Use the copy button.
- If you adjust or repair, run the points again and enter them as left.
When a point fails as found but passes as left, the overall result is not a plain pass. The sheet reports it as out of tolerance as found, so the owner knows to check what the instrument measured since its last calibration.
Decision rules and measurement uncertainty
Simple acceptance judges the error alone. Guarded acceptance passes a point only when the error plus your expanded uncertainty is still inside the limit, which lowers the risk of passing an instrument that is really out of tolerance. The non-binary rule accepts like simple acceptance and adds a conditional pass or conditional fail where the uncertainty overlaps the limit. ISO/IEC 17025 asks you to agree the rule with your customer before the work.